Recommissioning Dental Decontamination Equipment

Recommissioning Dental Decontamination Equipment

Share This Post

The 2020 outbreak of COVID-19 has caused the cessation of general dentistry, with the exception of emergency hubs. Long term closure causes several issues within dental practices, especially with decontamination equipment. In this first in a series of posts we will look at steps that can be taken for the recommissioning dental decontamination equipment; Washer Disinfectors, Autoclaves and RO Units.

Recommissioning Dental Decontamination Equipment – Ultrasonics

Clean the tank of the ultrasonic as well as the externals. Perform any manufacturer required validation tests, such as protein tests and soil tests. This is also a good time to check if the detergent you use is in date. In addition it is worth checking that the detergent you use is effective against enveloped viruses such as COVID-19.

Clean tank and externals.
50%
Validation checks performed and passing.
Ready to go! 100%
Recommissioning Washer Disinfectors

Recommissioning Washer Disinfectors

Clean the chamber fully, removing any dust from the internal surfaces and door seal. Ensure the water and power are on and run the washer on a full cycle using water instead of detergent. Replace the detergent and run again to ensure the detergent is now in place throughout the inlet tubes. Complete validation measures including Wash Checks, Residual Protein and Thermometric tests to confirm that processes are running as you had intended. Update your recommissioning documents, logbooks and reorder any detergents or validation as required.  

Clean Chamber, Door Seal and Internal Surfaces.
25%
Power and water on. Perform Full Cycle using water only.
50%
Fresh detergent added and cycle performed.
75%
Validation checks performed and passing.
Ready to go! 100%

Recommissioning RO Units

 

Fill and drain the unit at least twice before using any water. Test water quality using a MHPC Red Sampler and take a TVC reading. Do not use the RO water until it is suitable for the purpose. If being used for your DUWL make sure it is less then 100 CFU per ml. Use purchased water until the results are suitable. Contact your service provider if replacement filters or additional cleaning measures are required.

Fill and Drain RO Unit twice.
33%
Perform a MHPC Sampler Test. Ensure the test is passing before use.
66%
Filters changed or additional cleaning if required.
Ready to go! 100%

Recommissioning Autoclaves

 

Contact your service provider for support and recommissioning steps. These may include cleaning the internal surfaces and chamber including the door seal, cleaning the reservoir and flushing with suitable water (fill and drain) before refilling and running the validation cycles and tests. Update your recommissioning documents, logbooks and reorder any validation tests as required.

Contact service provider for support.
20%
Clean internal surfaces and door seal.
40%
Clean reservoir, fill and drain with suitable water.
60%
Run validation cycles and tests.
80%
Update recommissioning documents and logbooks.
Ready to go! 100%

Please get in touch with us if you need any assistance in preparation for practice reopening. In our next insight we will look at instrument reprocessing in practice.

Subscribe to our newsletter

The latest infection control updates, special offers and a lot more!

More News

Uncategorised

The Decontamination Cycle – Disinfection

In the dental decontamination cycle, disinfection is a bridging step between cleaning and sterilisation. It is commonly misunderstood and often conflated with cleaning – they

Handpiece Care

How Dental Handpieces Become Contaminated

Dental Handpieces are a critical tool used in routine and complex dental treatments. Inside the head of a high-speed handpiece, a turbine the size of